1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
CALCulate<prb>:CONVert:PARameter:VALue? <param>
Return the value of the specified conversion parameter for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
The <param> parameter is the keyword of the parameter as listed in Table 21 on page
74.
CALCulate<prb>:CONVert:PARameter:VALue <param>,<num>
Set the value of the specified conversion parameter for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
The <param> parameter is the keyword of the parameter as listed in Table 21 on page
74.
The <num> parameter is the new value with type and range per the list in Table 21 on
page 74.
To activate the new values for the parameters MINOP and MAXOP and update the
display, the command CALC:CONV:UPD must be sent, the probe must be removed
and inserted, or the unit power must be cycled off and on.
NOTE: This command is protected, which requires a password to set it.
Error '-2003, "Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN
CALCulate<prb>:CONVert:SNUMber?
Return the probe serial number for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
CALCulate<prb>:CONVert:SNUMber <seri>
Set the probe serial number for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
52
Shop for Fluke products online at:
Example: CALC2:CONV:PAR:CAT?
Response: "R0", "A", "B", "C", "MINOP", "MAXOP"
Example: CALC2:CONV:PAR:VAL? RTPW
Response: 100.0145
Example: CALC2:CONV:PAR:VAL RTPW,100.0145
Example: CALC2:CONV:SNUM?
Response: A_336C