Fluke 1523 Technical Manual page 64

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

CALibrate:DEVice:RANGe:NUMber <val>
Set the number of calibration points to use for the preselected range.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
The parameter <val> must be 2 to 4.
NOTE: This command is protected, which requires a password to set it.
Error '-203,"Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN.
CALibrate:DEVice:RANGe:REFerence<num>?
Return the reference value for a calibration point of the preselected range.
The REFerence suffix, <num>, specifies the calibration point, 1 to the number set with
the associated command CAL:DEV:RANG:NUM. If omitted, reference values for all
points are returned, separated by commas.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
The response is expressed in ohms or millivolts depending on the range being
calibrated.
CALibrate:DEVice:RANGe:REFerence<num> <val>
Set the reference value for a calibration point of the preselected range.
The REFerence suffix, <num>, specifies the calibration point, 1 to the number set with
the associated command CAL:DEV:RANG:NUM. Defaults to 1 if omitted.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
The parameter <val> is the reference value expressed in ohms or millivolts depending
on the range being calibrated.
The reference points, 1 to the specified number, must be entered in ascending order.
NOTE: This command is protected, which requires a password to set it.
Shop for Fluke products online at:
Example: CAL:DEV:RANG:NUM?
Response: 2
Example: CAL:DEV:RANG:NUM 2
Example: CAL:DEV:RANG:REF2?
Response: 10.568
Example: CAL:DEV:RANG:REF2 10.568
MyFlukeStore
www.
Digital Communication Interface
Serial Commands - Alphabetic Listing
1.877.766.5412
.com
57