1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
The response is expressed in ohms or millivolts depending on the range being
calibrated.
CALibrate:DEVice:RANGe:ADJust<num> <val>
Set the calibration offset from the reference value for a calibration point of the prese-
lected range.
The ADJust suffix, <num>, specifies the calibration point, 1 to the number set with the
associated command CAL:DEV:RANG:NUM. Defaults to 1 if omitted.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
The parameter <val> is the offset value expressed in ohms or millivolts depending on
the range being calibrated.
NOTE: This command is protected, which requires a password to set it.
Error '-203,"Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN.
CALibrate<chn>:DEVice:RANGe:CLEar
Reset all calibration points for the preselected range of the specified channel to
defaults.
The CALibrate suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
All calibration points are set to a gain of one and an offset of zero.
NOTE: This command is protected, which requires a password to set it.
Error '-203,"Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN.
CALibrate:DEVice:RANGe:NUMber?
Return the number of calibration points for the preselected range.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
The response is the number of calibration points in use.
56
Shop for Fluke products online at:
Example: CAL:DEV:RANG:ADJ2?
Response: .107
Example: CAL:DEV:RANG:ADJ2 -.065
Example: CAL:DEV:RANG:CLE
MyFlukeStore
www.
1.877.766.5412
.com