Serial Commands - Alphabetic Listing - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer

Serial Commands - Alphabetic Listing

Functions or
SCREEN
Groups
PARAMETER
System -
BAUD RATE
Communication
BAUD RATE
SERIAL PORT
System -
(all)
Information
(none)
(none)
(none)
(none)
DATE
DATE
TIME
TIME
System - Password
(none)
PASSWORD
(none)
USER PASSWORD
*indicates the command is password controlled. Reference "System - Password" section for password input.
6.4 Serial Commands - Alphabetic Listing
Each command description provides the structure (long and short format), a descrip-
tion of the command purpose, a command example, an example of what the command
returns (as applicable to query commands), and notes specific to the command. The
following apply to each group of commands:
48
Shop for Fluke products online at:
Command
SYST:COMM:SER:BAUD?
SYST:COMM:SER:BAUD <baud>
SYST:COMM:SER:OFF
*IDN?
*CLS
SYST:SNUM?
SYST:VERS?
SYST:ERR?
SYST:DATE?
SYST:DATE <year>,<month>,<day>
SYST:TIME?
SYST:TIME <hour>,<minute>,<second>
SYST:PASS:CDIS
SYST:PASS:CEN
SYST:PASS:CEN:STAT?
SYST:PASS:NEW
Numeric data, specified by the mnemonic, <num>, uses ASCII characters to
represent numbers. Numbers may contain a plus or minus ('+' or '-') sign,
decimal point ('.'), and exponent ('E' or 'e') with its sign. If a fractional
component is received when only an integer is required, the number is rounded
to the nearest integer without any resulting error message. The mnemonics DEF,
MIN, and MAX are often acceptable for the default, minimum, and maximum
value respectively. Unit suffixes, such as V or OHM, can be appended to
numeric parameters and are accepted without error but ignored.
Unrecognized commands or commands with incorrect syntax or invalid
parameters generate error messages in the error queue.
Upper case letters designate syntax that is required when issuing the command.
Lower case letters are optional and may be omitted.
< > indicates a required parameter.
[ ] indicates optional parameters.
( ) indicates a group of parameters that must be used together.
MyFlukeStore
www.
Password
Protection
Group
Read/Write
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
conditional
1.877.766.5412
.com
R
W
W
R
W
R
R
R
R
W
R
W
W
W
R
W

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1524

Table of Contents