Fluke 1523 Technical Manual page 62

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

CALibrate<chn>:DEVice:LOCK<bool>
Exclusively locks the channel to the inserted probe, or unlocks either.
The CALibrate suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
The parameter <bool> is 1 to exclusively lock the channel to the inserted probe. No
lock takes place and error '-200, "Execution error"' is queued if no probe is inserted, if
the channel is already locked, or if the probe is already locked.
The parameter <bool> is 0 to unlock the channel, the inserted probe if any, or both.
If no probe is inserted, only the channel is unlocked. If either is already unlocked, no
action is taken for the unlocked item and no error is queued.
NOTE: This command is protected, which requires a password to set it.
Error '-203,"Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN.
CALibrate<chn>:DEVice:LOCK:DATA?
Returns the lock details of the channel and the inserted probe if any.
The CALibrate suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
The first item returned is the probe serial number to which the channel is exclusively
locked. If the channel is not locked, "" is returned.
The second item returned is the unit serial number to which the probe is exclusively
locked. If the probe is not locked or no probe is inserted, "" is returned.
The third item returned is the unit channel to which the probe is exclusively locked. If
the probe is not locked or no probe is inserted, 0 is returned.
CALibrate:DEVice:RANGe:ADJust<num>?
Return the calibration offset from the reference value for a calibration point of the
preselected range.
The ADJust suffix, <num>, specifies the calibration point, 1 to the number set with the
associated command CAL:DEV:RANG:NUM. If omitted, offset values for all points
are returned, separated by commas.
The range must have been preselected with the command
CAL<chn>:DEV:RANG:SEL.
Shop for Fluke products online at:
Example: CAL2:DEV:LOCK?
Response: 1
Example: CAL2:DEV:LOCK 1
Example: CAL2:DEV:LOCK:DATA?
Response: "ALPHA_5","51789",2