Fluke 1523 Technical Manual page 73

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
MEASure? <prb>
Return the last measurement for the specified probe, with units according to the
UNIT:TEMP setting if it is a temperature.
The <prb> parameter specifies the probe, 1 or 2. Defaults to 1 if omitted.
If there is no valid measurement available the response is "0.0,OL".
READ? <prb>
Return the last measurement for the specified probe, with units according to the
UNIT:TEMP setting if it is a temperature.
The <prb> parameter specifies the probe, 1 or 2. Defaults to 1 if omitted.
If there is no valid measurement available the response is "0.0,OL".
SENSe<prb>:DATA:MV?
Return the present direct or thermocouple millivolt reading for the specified probe.
The SENSe suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
The response is expressed in millivolts. If there is no valid measurement available the
response is "0.0,OL".
If the presently selected input type is not millivolts or thermocouple, there will be no
response and error '-221,"Settings conflict"' will be queued.
SENSe<prb>:DATA:OHMS?
Return the present direct, PRT, RJ, or thermistor ohms reading for the specified probe.
The SENSe suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
The response is expressed in ohms. For a thermocouple probe, the response is the
reference junction ohms value. If there is no valid measurement available the response
is "0.0,OL".
If the presently selected input type is millivolts, there will be no response and error
'-221,"Settings conflict"' will be queued.
66
Shop for Fluke products online at:
Example: LOG:LAB18:NAME BATH
Example: MEAS? 1
Response: 0.127
Example: READ? 2
Response: 55.011
Example: SENS2:DATA:MV?
Response: 7.356895
MyFlukeStore
www.
1.877.766.5412
.com

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1524

Table of Contents