1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
NOTE: This command is protected, which requires a password to set it.
Error '-203,"Command protected"' is queued if the old password has not been entered
with SYST:PASS:CEN.
SYSTem:SNUMber?
Return the device serial number.
If the serial number has not been set at the factory, a 0 is returned.
SYSTem:TIME?
Return the current system time in 24 hour format.
The response is returned in the 24 hour format, <hour>,<minute>,<second>.
This command is only available in the model 1524, and results in a command error if
used with the model1523.
SYSTem:TIME <hour>,<minute>,<second>
Set the current system time.
The <hour>, <minute>, and <second> parameters must all be supplied.
The <hour> parameter is a one or two-digit number, 0 to 23.
The <minute> parameter is a one or two-digit number, 0 to 59.
The <second> parameter is a one or two-digit number 0 to 59.
This command is only available in the model 1524, and results in a command error if
used with the model1523.
SYSTem:VERSion?
Return the SCPI version number.
72
Shop for Fluke products online at:
Example: SYST:PASS:NEW 4321
This command has no response.
Example: SYST:SNUM?
Response: 509002
Example: SYST:TIME?
Response: 18,43,23
Example: SYST:TIME 18,43,23