Fluke 1523 Technical Manual page 76

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

bits 13 to 14:
STATus:QUEStionable:CONDition?
Return the Questionable Condition Register, indicating any measurement warning
conditions.
The bit assignments are the same as for the STATus:QUEStionable?.
STATus:QUEStionable:ENABle?
Return the Questionable Event Enable Register.
The bit assignments are the same as for STATus:QUEStionable?
STATus:QUEStionable:ENABle <num>
Set the Questionable Event Enable Register.
The <num> parameter specifies the event enable bits which are to be set, 0 to 65535
decimal. The bit assignments are the same as for the STATus:QUEStionable?.
SYSTem:COMMunicate:SERial:BAUD?
Return the present serial port baud rate setting.
The response is the present baud rate setting.
Shop for Fluke products online at:
bit 4:
probe 1 locked
bits 5 to 7:
not used
bit 8:
probe 2 not inserted
bit 9:
probe 2 measurement under range
bit 10:
probe 2 measurement over range
bit 11:
probe 2 EEPROM fault (bad checksum or
version number is not supported)
bit 12:
probe 2 locked
not used
bit 15:
Data logging EEPROM initialization failure
Example: STAT:QUES?
Response: 256
Example: STAT:QUES:COND?
Response: 256
Example: STAT:QUES:ENAB?
Response: 0
Example: STAT:QUES:ENAB 1799
MyFlukeStore
www.
Digital Communication Interface
Serial Commands - Alphabetic Listing
1.877.766.5412
.com
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1524