1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
LOGging:AUTomatic:LABel <num>
Set the tag number to which the LOG:AUT:POIN, LOG:AUT:STAT, and
LOG:AUT:VAL commands presently apply.
The <num> parameter specifies tag number, 1 to 25.
Error '-221,"Settings conflict"' is queued if the data logging storage is corrupt.
LOGging:AUTomatic:POINt?
Return the number of Auto-Log entries already stored for a preselected tag number.
The tag number addressed must have been previously selected with the
LOG:AUT:LAB command.
Error '-221,"Settings conflict"' is queued if the data logging storage is corrupt, or if no
tag is selected with LOG:AUT:LAB.
LOGging:AUTomatic:PRINt <num>
Send all Auto-Log entries for a tag number.
The <num> parameter specifies tag number, 1 to 25.
Each data line includes the default or user defined tag name, the channel number, the
primary value and units, and the time and date. When both channels are active, records
for each channel alternate. When only one channel is active, no records are sent for the
inactive channel.
Error '-221,"Settings conflict"' is queued if the data logging storage is corrupt.
60
Shop for Fluke products online at:
Example: LOG:AUT:LAB?
Response: 15
Example: LOG:AUT:LAB 4
Example: LOG:AUT:POIN?
Response: 28