1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
Error '-203,"Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN.
CALibrate<chn>:DEVice:RANGe:SELection?
Return the range presently being calibrated for the specified channel.
The CALibrate suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
The response is one of the keywords listed in Table 22 on page 75.
CALibrate<chn>:DEVice:RANGe:SELection <val>
Set the range presently being calibrated for the specified channel.
The CALibrate suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
The parameter <val> is one of the keywords in Table 22 on page 75. The value NOR-
MAL exits calibration mode, reconnects the inserted probe if any, and refreshes the
display.
This command sets the range to which all of the following commands apply:
CAL:DEV:RANG:CLEA
CAL:DEV:RANG:NUM
CAL:DEV:RANG:REF
CAL:DEV:RANG:ADJ
DISPlay<chn>:RESolution?
Return the maximum display resolution setting for the channel.
The DISPlay suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
DISPlay<chn>:RESolution <num>
Set the maximum display resolution for the channel.
The DISPlay suffix, <chn>, specifies the channel, 1 or 2. Defaults to 1 if omitted.
58
Shop for Fluke products online at:
Example: CAL:DEV:RANG:SEL?
Response: LO_OHMS
Example: CAL:DEV:RANG:SEL LO_OHMS
Example: DISP1:RES?
Response: 2
MyFlukeStore
www.
1.877.766.5412
.com