Fluke 1523 Technical Manual page 66

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

FETCh? <prb>
Return the last measurement for the specified probe, with units according to the
UNIT:TEMP setting if it is a temperature.
The <prb> parameter specifies the probe, 1 or 2. Defaults to 1 if omitted.
If there is no valid measurement available, the response is "0.0, OL".
LOGging:AUTomatic:DELete <num>
Delete all entries from an auto tag number.
The parameter <num> is the tag number for which all entries are to be deleted, 1 to 25,
or 0 to delete all entries for all tags.
Error '-221,"Settings conflict"' is queued if the data logging storage is corrupt, or if
the automatic delete or send display is active.
LOGging:AUTomatic:FREE?
Return the number of Auto-Log entries free, and in use by all tags.
The response consists of the number free entries, followed by the number of used
entries, separated by a comma.
Error '-221,"Settings conflict"' is queued if the data logging storage is corrupt.
LOGging:AUTomatic:LABel?
Return the tag number to which the LOG:AUT:POIN, LOG:AUT:STAT, and
LOG:AUT:VAL commands presently apply.
Return a value of 1 to 25, or 0 if not set.
Error '-221,"Settings conflict"' is queued if the data logging storage is corrupt.
Shop for Fluke products online at:
Example: DISP1:RES 3
Example: FETC? 1
Response: 105.236
Example: LOG:AUT:DEL 10
Example: LOG:AUT:FREE?
Response: 3760,1126
MyFlukeStore
www.
Digital Communication Interface
Serial Commands - Alphabetic Listing
1.877.766.5412
.com
59