Fluke 1523 Technical Manual page 74

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

SENSe:DATA:RJ?
Return the present internal or external reference junction temperature reading for the
specified probe.
The response is expressed in degrees Celsius.
If the presently selected input type is not millivolts or thermocouple, there will be no
response and error '-221,"Settings conflict"' will be queued.
SENSe:RJ:STATe?
Return the state of the reference junction for a given probe.
The response is OFF (no compensation), ON (automatic internal compensation), or
EXT (external compensation). If EXT is selected, the SENS:RJ:TEMP command must
be used to set the external compensation temperature.
SENSe:RJ:TEMPerature?
Return the external reference junction temperature for a given probe.
The response is expressed in degrees Celsius.
SENSe:RJ:TEMP <val>
Set the external reference junction temperature for a given probe.
The parameter <val> is expressed in degrees Celsius, and may range from -10C to
60C.
STATus:MEASure:EVENt?
Read and clear the Measurement Event Register, indicating whether or not a new mea-
surement is available to be read.
Shop for Fluke products online at:
Example: SENS2:DATA:OHMS?
Response: 100.45
Example: SENS:DATA:RJ?
Response: 23.568
Example: SENS:RJ:STAT?
Response: EXT
Example: SENS:RJ:TEMP?
Response: 22.731
Example: SENS:RJ:TEMP 23.123