1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
SYSTem:COMMunicate:SERial:BAUD <baud>
Set the serial port baud rate.
The <baud> parameter is the new baud rate, 2400 or 9600.
After this command is sent, the remote program must switch to the same baud rate for
all commands sent to the unit.
SYSTem:COMMunicate:SERial:OFF
Turn off the serial port.
After this command is sent, the unit will not respond to any further commands. The
port control item on the Instrument Setup screen of the unit must be used to turn the
serial port back on.
SYSTem:DATE?
Return the date.
The response is returned in the format <year>,<month>,<day>.
This command is only available in the model 1524, and results in a command error if
used with the model1523.
SYSTem:DATE <year>,<month>,<day>
Set the date.
The <year> parameter is a four-digit number, 2000 to 2099.
The <month> parameter is a one or two-digit number, 1 to 12.
The <day> parameter is a one or two-digit number. 1 to 31.
This command is only available in the model 1524, and results in a command error if
used with the model1523.
70
Shop for Fluke products online at:
Example: SYST:COMM:SER:BAUD?
Response: 9600
Example: SYST:COMM:SER:BAUD 2400
Example: SYST:COMM:SER:OFF
Example: SYST:DATE?
Response: 2000,9,22