The <month> parameter is a one or two-digit number, 1 to 12.
The <day> parameter is a one or two digit number, 1 to 31.
NOTE: This command is protected, which requires a password to set it.
Error '-2003, "Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN
CALCulate<prb>:CONVert:NAMe?
Return the keyword of the selected conversion type for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
Available conversion types are listed in Table 20 on page 73.
CALCulate<prb>:CONVert:NAM <conv>
Set the conversion type for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
The <conv> parameter is the keyword for the conversion type as listed in Table 20 on
page 73.
To activate the new probe conversion type and update the display, the command
CALC:CONV:UPD must be sent, the probe must be removed and inserted, or the unit
power must be cycled on and off.
NOTE: This command is protected, which requires a password to set it.
Error '-2003, "Command protected"' is queued if the password has not been entered
with SYST:PASS:CEN
CALCulate<prb>:CONVert:PARameter:CATalog?
Return the keywords of the probe characterization parameters associated with the cur-
rent conversion type for the specified probe.
The CALCulate suffix, <prb>, specifies the probe, 1 or 2. Defaults to 1 if omitted.
Each item in the response list is enclosed in double-quotes and separated from other
items by a comma. Each item indicates the keyword of a characterization parameter
for the current conversion type. If there are no parameters available for the current
conversion type, an empty string is returned.
The list of parameters depends on the selected conversion type as listed in Table 21 on
page 74.
Shop for Fluke products online at:
Example: CALC1:CONV:DATE:CAL 2000,8,29
Example: CALC1:CONV:NAM?
Response: CVD