Table 31 1523/24 Alignment Settings; Alignment - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

79. Set the output of the voltage source to 75 mV.
80. Query the reading every 2 seconds for 40 readings using the following
7.9.1.3.6
Reference Junction Ohms Range
81. Set the UUT in the RJ_OHMS cal range by sending the following commands
82. Connect a 10K Ohm resistor to the calibration cable connected to the UUT.
83. Query the reading every 2 seconds for 40 readings using the following
84. For all As Found data, calculate the average and standard deviation for each
7.9.2

1523/24 Alignment

Prior to performing the alignment, the following UUT configuration and control set-
tings must be set to the values indicated in the 1523/24 Alignment Settings table.
Caution: Failure to do this will cause large errors in the measurement readings
and invalidate the calibration.

Table 31 1523/24 Alignment Settings

Setting
NUM
NUM
NUM
NUM
NUM
Ref1, Ref2
Ref1, Ref2
Ref1, Ref2
Shop for Fluke products online at:
Calibration of Your Reference Thermometer Readout
command:
MEAS?
Record the data.
through the serial port:
CAL1:DEV:RANG:SEL RJ_OHMS
NOTE: Only connect pins 1and 4 to the 10K Reference resistor. This is a 2 wire
test.
command:
MEAS?
Record the data.
set of data. Compare the standard deviation to the specification. If the UUT
does not meet the standard deviation specification, re-take data at the suspect
range(s). If the data is still exceeds the specification, contact an Authorized
Service Center for assistance. Verify the average of each data set is less than the
Accuracy limit listed in the Accuracy Test Settings and Specification tables.
Cal Range
MV
L75_OHMS
LO_OHMS
MED_OHMS
HI_OHMS
MV
L75_OHMS
LO_OHMS
MyFlukeStore
www.
Value
Description
2
MV Range # of cal points
2
L75_OHMS Range # of cal points
2
LO_OHMS Range # of cal points
2
MED_OHMS Range # of cal points
2
HI_OHMS Range # of cal points
0, 75
mV Ref Calibration points
0, 75
L75_OHMS Ref Calibration points
75, 400
LO_OHMS Ref Calibration points
Manual Calibration Process
1.877.766.5412
.com
89

Hide quick links: