4 Operation

4.1 Connect the power supply unit

A
C
12
4
Operation
Operation
4.1
Connect the power supply unit
Connect the power supply unit
IMPORTANT
Check that the power supply voltage corresponds to the voltage data on
the power supply unit before connecting the power supply cable.
1. Snap the power cable connector into the connecting socket, on the
bottom side of the charging station.
2. Lay the cable into the channel adjacent to the connecting socket and
clamp the cable in under the holders.
IMPORTANT
Position the device at least 1.5 m away from the patient on a stable,
horizontal surface. Position the unit so it is easy to disconnect it from the
power supply.
➢ Pull out the plug to disconnect the unit from the power supply.
4.2

Replace and charge the battery

The battery can operate continuously for up to 25 minutes. Once the
Replace and charge the battery
battery has been completely discharged, a full recharge takes approx. 1
hour. The light pen starts to blink when the battery is running low.
The batteries can be stored for 4 months. Once this period has expired,
the battery must be charged. If this is not the case, it may cause the
battery to have reduced charging capacity or to fail completely. Sirona
B
recommends storing the battery on the charging station.
As a general rule, the batteries included with the delivery have a service
period of up to 12 months. After this period, power decreases accordingly.
Please check the usage period based on the LOT number (definition of
LOT = week week year year e.g. 0212 for calendar week 2, 2012).
CAUTION
Risk of crushing
When sliding open the battery, the skin on your fingers may become
trapped.
➢ Hold the battery as far back as possible.
1. Remove the discharged battery (B) from SIROInspect (A).
2. Insert the charged battery into SIROInspect until it audibly clicks into
place.
3. Position the discharged battery onto the charging station (C) until it
audibly clicks into place.
 As soon as the charging light switches on, the battery is fully
charged.