MACH Pt
In aeronautical units (ft/kts or m/km/h), this key selects the Mach display.
In pressure units (mbar, inHg etc.), this key selects the total pressure (Pt) display.
If the altitude aim is changed, after the entry of CAS, the Mach aim value changes while the
CAS value remains unchanged.
If the altitude aim is changed, after the entry of Mach, the CAS aim value changes while the
Mach value remains unchanged.
If the Ps aim is changed, after the entry of Qc, then the Pt aim value changes while the Qc
remains unchanged.
If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the Pt
remains unchanged.
When in Control Mode (AIM display), a new Mach or Pt aim can be entered using the numeric
keys, Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.

ROC Ps RATE

In aeronautical units (ft or m), this key selects the Rate of Climb (ROC) or vertical speed
display.
In pressure units (mbar, inHg etc.), this key selects the rate of change of Ps display.
Select [UNITS], from the main menu, to change between aeronautical and pressure units.
When in Control Mode (AIM display), a new ROC or Ps rate can be entered using the numeric
keys. Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.
The default control rate set at manufacture is 3000 ft/min.
6 - 3