Druck ADTS 505 User Manual
MACH Pt
•
In aeronautical units (ft/kts or m/km/h), this key selects the Mach display.
•
In pressure units (mbar, inHg etc.), this key selects the total pressure (Pt) display.
•
If the altitude aim is changed, after the entry of CAS, the Mach aim value changes while the
CAS value remains unchanged.
•
If the altitude aim is changed, after the entry of Mach, the CAS aim value changes while the
Mach value remains unchanged.
•
If the Ps aim is changed, after the entry of Qc, then the Pt aim value changes while the Qc
remains unchanged.
•
If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the Pt
remains unchanged.
•
When in Control Mode (AIM display), a new Mach or Pt aim can be entered using the numeric
keys, Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.
ROC Ps RATE
•
In aeronautical units (ft or m), this key selects the Rate of Climb (ROC) or vertical speed
display.
•
In pressure units (mbar, inHg etc.), this key selects the rate of change of Ps display.
•
Select [UNITS], from the main menu, to change between aeronautical and pressure units.
•
When in Control Mode (AIM display), a new ROC or Ps rate can be entered using the numeric
keys. Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.
The default control rate set at manufacture is 3000 ft/min.
6 - 3
6
K0260 Issue No. 8