GE Druck ADTS 505 User Manual page 108

Data test set
Table of Contents

Advertisement

6 - 28
2.
Use the Up and Down keys to highlight the required aeronautical and pressure units.
3.
Press SAVE SETTINGS to store in non-volatile memory or press MAIN MENU to store in volatile memory
and return to the Main Menu Screen. Press RESTORE DEFAULTS to restore the last set of stored values
and unit selection saved as a default.
Select limits
Enter the PIN to access the SETUP LIMITS
display. Select one of the preset limits of MAX,
STANDARD and CIVIL or define and enable USER
1 limits. An eight character name/code can
replace USER 1 title.
Default limits
This set of limits, identified as: CIVIL are enabled
when the ADTS 505 is switched on:
MIN ALT
MAX ALT
MIN CAS
MAX CAS
MAX MACH
MAX ROC
ARINC LIMITS
ALT CORRECTION
MIN Ps
115.972 mbar (2.7 inHg)
MAX Ps
1050.406 mbar (38.1000 inHg)
MIN Qc
0.0 mbar
MAX Qc
368.01 mbar
MAX Rate Ps 109.85 mbar/min (10.8127 inHg/min)
MAX Rate Qc 109.85 mbar/min (10.8127 inHg/min)
K0260 Issue No. 8
-1000 ft
50,000 ft
0.0 knots
450 knots
1.0 Mach
6,000 ft/min
OFF
0 ft
(10.8673 inHg)
Reference and Specification