Additional Messages
Internal temperature
The temperature sensor in the pressure measurement system measures the temperature of
the pressure manifold. At a temperature greater than 60°C the sensor triggers an overheat
fault. In measure mode, the display flashes between the normal aim/measure screen and
"Ps Sensor Outside Calibrated Range". In control mode, the display shows "CONTROL
HALTED" the system stops controlling and switches off the pumps. When the sensor detects
a temperature of 55°C the display stops flashing.
Before re-entering control mode check the air vents for obstructions. To re-enter control
mode press LEAK MEASURE/CONTROL twice. If this overheat fault happens again switch off
the unit and return to repair depot.
5 - 17
5
K0260 Issue No. 8