Download  Print this page

GE Druck ADTS 505 User Manual Page 117

Data test set.
Hide thumbs

Advertisement

Druck ADTS 505 User Manual
6 - 37
DK286 – Touch Screen Calibrator
This program enables the calibration of the touch screen. To execute this application, exit from the ADTS 505
program and double click on the Calibrate Touch icon
. The Stylus application, stored as part of the
system application, executes first. Select the calibration option and click the Recalibrate button. Follow the
on-screen instructions, after obtaining all the calibration points, touch the screen once and the Stylus
program returns to the previous screen. Close down the Stylus application, the screen shows the following:
This screen gives the opportunity to save the calibration by clicking the Save Screen Calibration button, or
restoring the default screen calibration stored by clicking the Restore Default Calibration Data button. The
system stores the calibration data.
On completion, click the Exit button.
DK287 – Display Backlight Brightness Setting Program
This program enables the user to adjust the brightness of the display back-light. To execute this program, exit
the ADTS205 program and double click on the icon Set BkLight Brightness
. The screen shows the
following:
6
The slide bar Brightness Level starts in the position of the current back-light brightness. Adjust the brightness
level by moving the slide bar left for darker and right for brighter. The back-light changes as the slide bar
moves. To save the setting, click the Save Brightness Setting button, this stores the current brightness setting.
On completion, click the Exit button.