Performance expressed in pressure units
Units
(1)
71.7
Maximum Range
(2)
Accuracy
Repeatability
Resolution
(1)
Maximum operating range can be achieved using the internal pump at airfield altitudes
to a maximum of 7,500 ft above sea level.
(2)
The accuracy figures stated include 12 months stability.
(3)
The system measures below this value but, because of pump capacity, controls to 71.7
mbar absolute. 35 mbar is the lowest calibration point.
Max Pt 3000 mbar absolute (with Ps >500 mbar absolute)
Max Qc 2500 mbar differential (with Ps <500 mbar absolute)
Altitude Sensor
(Ps)
mbar abs
inHg abs
(3)
to 1355
2.1 to 40
±0.1
±0.003
±0.05
±0.0015
0.01
0.0003
Airspeed Sensor
(Qc)
mbar diff
-950 to +2490
±0.1% reading
±0.125 mbar
±0.05 rising to 0.17
±0.0015 rising to 0.0051
0.01
N
OTES
inHg diff
1 to 73.53
0.1 to 10
Better than
±0.1% reading
±0.0037 inHg
0.005
0.0003
0.001