6 - 2
ALT Ps
In aeronautical units (ft or m), this key selects an altitude display.
In pressure units (mbar, inHg etc.), this key selects a static pressure (Ps) display.
Select [Units], from the main menu, to change the display units.
Before entering a new altitude or Ps aim, set the rate of change using the ROC/RATEPs key.
When in Control Mode (Aim display), enter a new altitude or Ps aim using the numeric keys.
Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.

SPEED Qc

In aeronautical units (kts, km/h), this key selects an airspeed (CAS) display. In pressure units
(mbar, inHg etc.), this key selects dynamic pressure and the display shows (Qc).
Note: For testing purposes, Calibrated Airspeed (CAS) equates to Indicated Airspeed (IAS).
If the altitude aim is changed, after the entry of CAS, the Mach aim value changes while the
CAS value remains unchanged.
If the altitude aim is changed, after the entry of Mach, the CAS aim value changes while the
Mach value remains unchanged.
If the Ps aim is changed, after the entry of Qc, then the Pt aim value changes while the Qc
remains unchanged.
If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the Pt
remains unchanged.
Select [Units], from the main menu, to change the display units.
The rate of change is fixed at 300 kts/min or equivalent.
When in Control Mode (AIM display), a new CAS or Qc can be entered using the numeric keys.
Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.
K0260 Issue No. 8