GE Druck ADTS 505 User Manual page 42

Data test set
Table of Contents

Advertisement

3 - 14
To select a limit set (User)
Press SETUP, the display shows Setup, [Limits]
[xxxxxx], where xxxxxx is the current name of
the selected limit set. Select [Limits] [xxxx] F3.
The display highlights the current set of limits
in use by the name in the list. The current limit
values can be inspected by pressing [View
Details] F4. The display shows the current
limit values and divides these values into
aeronautical (Page #1) and pressure (Page
#2).
The ARINC 565 limit can be selected ON or
OFF in a set of user defined limits, When
selected, the values entered and/or achieved
in control mode are checked within the
operating envelope of the ARINC 565 limits,
see section 6.
To select another set of limits from the list
press [Select Limits], F5. Use [ ] or [ ] to
move through the list.
To edit a limit set (Supervisor)
To edit or make a new set of limits press [Enter
PIN], F3 and enter the PIN. Use [ ] or [ ] to
move through the list. Each set of limits is
identified by it's name. New sets of limits can
be created and existing sets of limits can be
renamed or values changed. Refer to section
6 for more details. The user-named sets, can
be edited by pressing [Select Limit [xxxxxx],
then Modify Details or Modify Name.
Select [Save Settings] to store the new details
or name to be used on the next power-up of
the system.
Notes:
1
For temporary changes press QUIT/CLEAR
in place of [Save Settings].
2
The factory-defined limit sets, "STANDARD",
"MAX" and "CIVIL" cannot be changed.
K0260 Issue No. 8
Limits
Active Limit Set: CIVIL
s
Available Preset Limits
t
MAX
STANDARD
Select
CIVIL
Limits
Available User Limits
View
Details
USER 1
USER 2
Enter
USER 3
PIN
USER 4
USER 5
7
8
ALT
SPEED
MACH
Ps
Qc
Pt
4
5
ROC
GROUND
RATE
RATE Ps
1
2
_
LEAK
MEASURE
HELP
SETUP
0
000
CONTROL
Min ALT
-1000 ft
Max ALT
50,000 ft
Min CAS
0.0 knots
Max CAS
450 knots
Max Mach
1.0 Mach
Max ROC
6,000 ft/min
Alt Cor.
0 ft
ARINC Limits
OFF
Page #1
7
ALT
SPEED
MACH
Ps
Qc
Pt
4
ROC
GROUND
RATE
RATE Ps
1
_
LEAK
MEASURE
HELP
SETUP
000
CONTROL
Min Ps
0.0 mbar
Max Ps
1050.406 mbar
Min Qc
0.0 mbar
Max Qc
368.01 mbar
Rt
Ps
105.00 mbar/min
Rt Qc
90.0 mbar/min
Page #2
7
ALT
SPEED
MACH
Ps
Qc
Pt
4
ROC
GROUND
RATE
RATE Ps
1
_
LEAK
MEASURE
HELP
SETUP
000
CONTROL
Operation
User
Supervisor
Limits
F1
s
F2
t
F3
Select
Limit
F4
Modify
Details
F5
Modify
Name
F6
Save
Settings
CLEAR
9
QUIT
6
ABORT
3
DELETE
.
ENTER
User
Supervisor
Details
Details
Next
Next
F1
Page
Page
s
F2
t
F3
Modify
F4
Units
ft kts
F5
ft/min
more
F6
(1 of 2)
CLEAR
8
9
QUIT
5
6
ABORT
2
3
DELETE
.
0
ENTER
User
Supervisor
Limits
Details
m km/h
Next
F1
m/min
Page
mbar
F2
mbar/min
F3
F4
F5
more
F6
(1 of 2)
CLEAR
8
9
QUIT
5
6
ABORT
2
3
DELETE
.
0
ENTER

Advertisement

Table of Contents
loading

Table of Contents