Go to GROUND
When the system is in control mode press:
Note: Go to ground does not operate in measure mode.
The ALT Aim value immediately changes to the current
ground pressure (local airfield altitude) and the system
safely controls to the GROUND value stored from
power-up. The CAS Aim value immediately changes to
zero and the system safely controls to zero:
The altitude and airspeed decreases at the rate set in
ROC and RtCAS.
When the altitude and airspeed are at ground and zero
the display changes to:
SAFE
AT
GROUND
7
ALT
SPEED
MACH
Ps
Qc
Pt
4
ROC
GROUND
RATE
RATE Ps
1
_
LEAK
MEASURE
HELP
SETUP
000
CONTROL
At ground and zero the display shows safe at ground for 5
F1
F2
seconds then the ADTS 505 system closes the output valves
F3
to isolate the aircraft system and goes into a system datum
F4
F5
check routine (includes operation of the pump and control
F6
valves) go to 3.7.8 for more details.
8
9
CLEAR
QUIT
5
6
ABORT
2
3
DELETE
.
0
ENTER
After completion of the system datum check routine the
display shows safe at ground and the message "Press Clear/
Quit to continue".
It is now safe to continue testing or to switch off and
disconnect the aircraft system.
When familiar with these procedures go to 3.5.4 to leak test
the ADTS 505.