Appendix A: Specifications

All specifications apply to the TPS2000 series oscilloscopes. P2220
probe specifications appear at the end of this chapter. To verify that
the oscilloscope meets specifications, the oscilloscope must first
meet the following conditions:
H The oscilloscope must have been operating continuously for
twenty minutes within the specified operating temperature.
H You must perform the Do Self Cal operation, accessible through
the Utility menu, if the operating temperature changes by more
than 5 °C (9 °F ).
H The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted "typical."
Oscilloscope Specifications
Acquisition
Acquisition Modes
Acquisition Rate, typical
Single Sequence
g
q
TPS2000 Series Digital Oscilloscope User Manual
Sample, Peak Detect, and Average
Up to 180 waveforms per second, per channel (Sample
acquisition mode, no measurements)
Acquisition Mode
Sample, Peak Detect
Average
Acquisition Stops After
Single acquisition, all channels
simultaneously
N acquisitions, all channels
simultaneously, N is selectable
from 4, 16, 64, and 128
A- 1