Holdoff. You can use the Trigger Holdoff function to produce a stable
display of complex waveforms, such as pulse trains. Holdoff is the
time between when the oscilloscope detects one trigger and when it
is ready to detect another. The oscilloscope will not trigger during
the holdoff time. For a pulse train, you can adjust the holdoff time so
the oscilloscope triggers only on the first pulse in the train.
Trigger level
Indicates
trigger points
Triggers are not recognized during holdoff time.
To use Trigger Holdoff, push the HORIZ MENU
option button and use the multipurpose knob to adjust the holdoff.
The resolution of the trigger holdoff varies depending on the
Horizontal SEC/DIV setting.
9- 46
Acquisition
interval
Holdoff
TPS2000 Series Digital Oscilloscope User Manual
Acquisition
interval
Holdoff
Holdoff
Acquisition
interval
Holdoff
Set Holdoff