Invacare Kite Service Manual page 24

Hide thumbs Also See for Kite:
Table of Contents

Advertisement

SERVICE MANUAL
PROBLEM
Mobility
device
judders in
drive mode
Batteries
not being
charged
Mobility
device runs
too slowly
Electrical
actuator
does not
respond
24
OTHER
POSSIBLE
SYMPTOMS
CAUSE
None
Batteries
defective
(unstable
voltage)
Drive motor(s)
defective
None
Batteries
defective
LEDs blinking
Charging unit
on charging
defective
unit
None
Remote
defective
Batteries
defective
Remote
Lighting/actuat
displays
or module
flashing "E",
faulty
status diode
on
light/actuator
module does
not extinguish
even if the
remote is shut
down or
disconnected
None
Cable
disconnected
or damaged
Electrical
actuator faulty
Remote faulty
SOLUTION
Documentation
Replace batteries
See chapter 8.4.
Replace motor(s)
See chapter 8.10.1.
Replace carbon
See chapter 8.10.4.
brushes
Replace batteries
See chapter 8.4.
Replace charging
See charging unit
unit
operating manual
Replace remote
See chapter 8.2.5.
Replace batteries
See chapter 8.4.
Replace
See chapter 8.1.5.
lighting/actuator
module
Ensure correct
See chapter 8.6.
cable
connection,
replace cable if
necessary
Check actuator
See chapter 7.3.
Replacing the
See chapter 8.2.5.
remote
®
Invacare
- Kite