Replacing The Connecting Cable Of The Operating Hours Counter - Invacare Kite Service Manual

Hide thumbs Also See for Kite:
Table of Contents

Advertisement

SERVICE MANUAL
9.3.2

Replacing the connecting cable of the operating hours counter

Requirements:
· Philipps screwdriver, size 2
· Side cutters
· Cable ties
·
The operating hours counter is located in the
side cover.
·
Disconnect the plug-in connection of the
operating hours counter from the drive
motor.
·
Remove the side cover, as described in
chapter 8.1.3.
·
Detach the operating hours counter from the
side cover.
·
Turn the operating hours counter around.
·
Loosen and remove the four Phillips screws
of the housing.
·
Disconnect the plugs (+) & (-).
·
Remove the cable ties (4).
·
Carefully pull the connecting cable (2) of the
operating hours counter out of the casing
(1).
·
Replace the cable of the operating hours
counter.