Operational Faults; Drive Fault Diagnosis - Invacare Kite Service Manual

Hide thumbs Also See for Kite:
Table of Contents

Advertisement

®
Invacare
- Kite
6

Operational faults

Different power modules can be installed in the mobility aid in connection with different remotes.
The rectification of operational faults depends on the installed power module in each case.
The power modules used are described in 4.2.
Note
The tables for rectification of operational faults listed in the following chapters are only an excerpt
from the original manufacturer's manuals.
You can obtain the original manuals from Invacare®.
If you have problems with the mobility aid, please proceed as follows:
· First assess the possible cause of the problem using the following table.
· Check the remote status display. Evaluate the error code.
· Carry out the necessary checks and repairs as recommended in the following table.
6.1

Drive fault diagnosis

PROBLEM
Mobility
device will
not start
OTHER
POSSIBLE
SYMPTOMS
CAUSE
The remote
Drive motors
status display
disengaged
illuminates
normally and
does not show
an error code.
Remote status
Batteries
display does
defective
not illuminate
Completely
discharge
battery
Power supply
to remote
interrupted
Remote
defective
Remote status
Various
display blinking
causes
SOLUTION
Documentation
See operating manual
Engage drive
motors
Replace batteries
See chapter 8.4.
Pre-charge
See operating manual
batteries
Check master
See chapter 8.5.
fuse
Check cables
See chapter 8.6.
between the
modules for loose
connections and
damage
Replace remote
See chapter 8.2.5.
Assess error code
See chapter 6.2.