Full Start Options

Setup for Internal Calibrator Use (Full Start)

VALVES: OFF
CAREFULLY FILL THE CALIBRATOR
W ITH A HALF CC OF CALIBRATION
ELECTROLYTE
--- PRESS ACK W HEN READY ---
1
2
VALVES: OFF
DO YOU W ANT TO CONNECT THE
OPTIONAL SPAN GAS TANK
AT THIS TIM E?
YES
1
2
VALVES: OFF
CONNECT A SAM PLE GAS
TO THE SAM PLE PORT
--- PRESS ACK W HEN READY ---
1
2
VALVES: PURGE-TO-SPAN
PURGE IN PROGRESS
TIM E REM AINING 00 : 01 : 00
PLEASE W AIT . . .
SKIP
1
2
ACK
ESC
3
4
5
NO
ESC
3
4
5
ACK
ESC
3
4
5
3
4
5
Teledyne Electronic Technologies
Analytical Instruments
Ultra
Ultra T T T T T race Oxyg
Ultra
race Oxyg
race Oxyg
Ultra
Ultra
race Oxygen Anal
race Oxyg
6. If you request the internal calibra-
tor, you are directed to fill the
calibrator with electrolyte. After
the process is completed, press
to continue or
ACK
the previous screen.
7. After pressing
ACK
next screen gives the option of
connecting the span gas tank.
Press
for internal calibrator
NO
only.
8. After pressing
in step 7, the
NO
system directs you to connect the
sample gas tank to the appropri-
ate inlet. After the process is
completed, press
ACK
sures that the sample gas is
connected to the sample inlet of
the analyzer
9. After pressing
ACK
system performs a sample line
purge, then automatically contin-
ues on to step 10.