Forcing a Cold Start Without First Shutting Down

To force a
COLD START
follow the procedure outlined below.
VALVES: ANALYZE
C=9:59
FLOW = 150.0 CC/MIN
5.00
PPM 02
AUTORANGE: 10 PPM
CAL
STDBY
SETUP
RANGE
1
2
3
VALVES: SHUTTING DOW N
TIM E REM AINING 00 : 02 : 00
PLEASE W AIT . . .
1
2
3
VALVES: STANDBY
STANDBY M ODE - ALARM S DISABLED
1
2
3
VALVES: OFF
SELECT THE BACKGROUND GAS
USING THE KNOB ON THE STATUS PANEL
1
2
3
VALVES: OFF
WARNING: IF THIS IS THE FIRST
SETUP THE FULL STARTUP
IS REQUIRED.
FULL
QUICK
1
2
3
Hold down the third key while
pressing the second (QUICK) key.
4-10
at the next power-up after the analyzer has been used,
1. From the Analyze mode main menu,
T=25C
press the
ALRMR
4
5
2. The next menu will indicate that the
system is beginning to shut down, and
indicates the time remaining until the
next screen appears.
4
5
3. From the Stand-by mode, press
to continue.
START
4
5
4. The menu will prompt you to select the
background gas from the status panel.
Press
menu.
ACK
ESC
4
5
5. Holding down the orange key while
pressing
choose a
reboot.
4
5
Teledyne Electronic Technologies
Analytical Instruments
key.
STDBY
to go to the first system
ESC
will allow you to
QUICK
for the next
COLD START
Chapter 4