Gas Line Connections

All of the gas lines to the system hook up at the back of the unit. See Figure 3-15.
All of the fittings are
compressed air inlet fitting. You have been provided with VCR glands, gaskets
and female nuts. The glands must be orbitally welded to electro-polished
tubing and then connected to the inlet/outlet connections on the instrument.
3
Use
/
" and
4
(P/N G284) between the fitting and tighten the female and male nuts until
finger-tight; then, by holding the male nut with the wrench, tighten the female
nut with the second wrench by 1/6 turn.
CAUTION:
Check that each of the lines is hooked up to the correct connection. The lines
should be connected in the following order:
1. Span Gas In
If you choose to use a span gas to calibrate the analyzer, connect the
span gas tank to the span gas inlet.
North Am erican
Jum per Bar
Fuse
Figure 3-14. Fuse Replacement
1
/
" male VCR-type fittings, with the exception of the
4
5
/
" wrenches to connect the gas lines. Insert the gasket
8
Do not put any torque on the tubes welded on the sam-
pling system.
Teledyne Electronic Technologies
Analytical Instruments
Ultra T T T T T race Oxyg
Ultra
Ultra
race Oxyg
race Oxyg
Ultra
Ultra
race Oxygen Anal
race Oxyg
Fuse Block
Cover
Jum per Bar
en Anal
en Anal
yz yzer er er er er
yz yz
en Anal
en Analyz
European 1.5 Am p Fuses
5 X 20 m m
1
/
"
4
3-19

Advertisement

Table of Contents
loading

Table of Contents