Teledyne 3060E Instruction Manual page 123

Ultra trace oxygen analyzer
Table of Contents

Advertisement

T T T T T r r r r r oub
oub
oub
leshooting and Maintenance
leshooting and Maintenance
oubleshooting and Maintenance
oub
leshooting and Maintenance
leshooting and Maintenance
PROCEED
PROCEED
NO SPAN GAS
GENERATED BY
INTERNAL CALIBRATOR
CHECK RESISTANCE ACROSS
CALIBRATOR ELECTRODES;
IT MUST BE SEVERAL K
IF RESISTANCE IS TOO LOW (FEW )
CALIBRATOR HAS INTERNAL SHORT
IF SO:
GENTLY PULL ELECTRODES
1
OUT OF WICK AT POINT
OF INSERTION
2
SEPARATE ELECTRODE WIRES
SO THEY DO NOT TOUCH
EACH OTHER
WET THE ENTIRE CALIBRATOR WICK
3
WITH CALIBRATION ELECTROLYTE
4
INSURE THERE IS ENOUGH
CALIBRATION ELECTROLYTE
IN THE CELL
5
MOUNT THE CALIBRATION ASSEMBLY
RECALIBRATE
ANALYZER
(SECTION 4.3.7
OR 4.3.7.1)
YES
SPAN GAS
GENERATED
REMOVE THE CALIBRATOR
1
ELECTROLYTE ASSEMBLY
2
REMOVE CALIBRATION ELECTROLYTE CUP
CLEAN AND DRY THE CUP AND THE INSIDE
3
OF CALIBRATOR MOUNTING BLOCK
REPLACE CUP IN THE MOUNTING BLOCK
4
5
ADD CALIBRATION ELECTROLYTE
6
WET THE ENTIRE CALIBRATOR WICK
WITH CALIBRATION ELECTROLYTE
7
MOUNT THE CALIBRATOR ASSEMBLY
RECALIBRATE
(SECTION 4.3.7
YES
SPAN GAS
GENERATED
Table 5-11. No Span Gas From Internal Calibrator
Teledyne Electronic Technologies
Analytical Instruments
Ultra T T T T T race Oxyg
Ultra
Ultra
race Oxyg
race Oxyg
Ultra
Ultra
race Oxygen Anal
race Oxyg
EXAMINE CALIBRATOR WICK
IF EXCESSIVELY COLORED,
RINSE WITH DISTILLED WATER
IF EXCESSIVE COLOR PERSISTS,
THE
NO
THE
ANALYZER
OR 4.3.7.1)
1
NO
2
en Anal
en Anal
yz yzer er er er er
yz yz
en Anal
en Analyz
REPLACE CALIBRATOR
REPLACE
CALIBRATOR
REPLACE
ELECTROLYTE
CUP
5-15