FLOW FAILURE
<20cc/min
150cc/MIN OPTIMUM
1
CHECK AIR SUPPLY
PRESSURE
(75-80 PSIG)
2
CHECK SAMPLE SPAN
GAS PRESSURE
(10-30 PSIG)
PRESS THE START KEY
TO RESET THE ANALYZER
TO THE FACTORY
SETTINGS
NO
FAILURE
CORRECTED
YES
PROCEED
5-10
NO FLOW INDICATED
CHECK MFC
1
CABLE &
CONNECTIONS
CHECK BACKGROUND
2
GAS SELECTION KNOB
ON FRONT PANEL IS SET
TO APPROPRIATE GAS
PRESS THE START KEY
TO RESET THE ANALYZER
TO THE FACTORY
SETTINGS
Table 5-6. Flow Failure Less than 20cc
Teledyne Electronic Technologies
Analytical Instruments
Chapter 5
Chapter 5
Chapter 5
Chapter 5
Chapter 5
NO
FAILURE
CORRECTED
YES
CALL
SERVICE