After the electrolyte has been purged, it can be added to the sensor using the
following procedure:
1. Remove the plastic screw cap from the anode port on the stainless
steel plate on the top of the sensor body (Figure 3-4.1) and flush out
the sensor by using oxygen-free gas.
Oxygen-Free
Gas In
Plastic/Teflon
Tube
Figure 3-4.1. Flushing Out the Ultra Trace Sensor by O
2. Insert the tube attached to the 50 cc syringe into the electrolyte
bottle and slowly fill the syringe with electrolyte.
3. Insert the tube attached to the syringe into the sensor through the
opening on the top stainless steel plate. Slowly pour electrolyte into
the sensor (Figure 3-4.2).
3-6
Gas Out
Teledyne Electronic Technologies
Analytical Instruments
Chapter 3
Chapter 3
Chapter 3
Chapter 3
Chapter 3
-Free Gas