Figure 4-7. Adjusting Span Sensitivity When Spiking Occurs on the Signal
VALVES: ZERO, INT SCRUB
OXYGEN= 5.0 PPM O2
FLOW = 150.0 CC/M IN
SETTLING DELAY = 299 SECS
W AIT FOR O2 SETTLING
OR ANY KEY TO ABORT
1
2
3
4-34
4. If you have chosen Auto settling, the
screen displays the present oxygen
concentration and the flow rate. It also
checks whether the background signal is
within the acceptable range.
TIME
Before the processor begins to check
for the slope of the baseline, a delay of 5
minutes occurs and is used to counter
any spikes that might occur during
mode switching. After an initial delay,
4
5
the screen displays the time left in
seconds while the system waits for
settling. After auto settling, the system
will automatically return you to the
Analyze mode main menu.