VALVES: SAMPLE PURGE
CHECK THE FLOW RATE FOR
150 CC/MIN
FLOW = 150 CC/MIN
--- PRESS ACK WHEN READY ---
1
2
3
FROM HERE GO TO
FROM HERE GO TO
FROM HERE GO TO
FROM HERE GO TO
FROM HERE GO TO
STEP 21
STEP 21
STEP 21
STEP 21
STEP 21
4-14
10. The system directs you to check
ACK
ESC
4
5
Teledyne Electronic Technologies
Analytical Instruments
the flow rate for 150 cc/min and
to press
when ready.
ACK
With recommended sample
gas pressure (10–30 psig), the
mass flow controller will
automatically control the flow
at 150 ± 4 sccm. You may
check the flow by connecting a
bubblometer at the analyzer
sample vent line to insure that
there is proper flow.