Adding Water To The Reservoir - Teledyne 3060E Instruction Manual

Ultra trace oxygen analyzer
Table of Contents

Advertisement

Model 3060E
Model 3060E
Model 3060E
Model 3060E
Model 3060E

Adding Water to the Reservoir

If the analyzer is to be used continuously for on-line gas analysis, then add
approximately 800 cc of distilled water to the reservoir. The rate of water
addition to the sensor is controlled by the metering valve MV3 located adjacent
to the reservoir, below the top cover plate. The valve MV3 is also accessible
from the rear of the analyzer.
The water drip rate is set at the factory at 1 drop every 2–4 minutes. This drip
rate is close to the rate of evaporation of water from the Ultra Trace sensor.
However, if the water drip rate is not proper, the electrolyte level in the Ultra
Trace sensor will not be maintained at the desired level (between the yellow and
green level markers). A drip rate that is too slow will cause the electrolyte level
to drop slowly. If this occurs, open MV3 slightly (
control knob anticlockwise. See Figure 3-11 for the location of MV3. Repeat
this step if necessary.
Figure 3-11. Piping Layout, Water Reservoir and Metering Valves
CAUTION:
If you open valve MV3 too wide, an excessive amount of water
will enter the sensor and the output will increase but will
slowly return to the original value. With MV3 open too wide,
the level detector will open and close more frequently causing
output instability. If this occurs, close MV3 slowly until the
drip rate is normal. If you experience difficulty in adjusting
MV3, you may add prepurged water directly to the sensor
through the filler port.
3-14
Teledyne Electronic Technologies
Analytical Instruments
1
/
turn) by rotating the valve
2
Chapter 3
Chapter 3
Chapter 3
Chapter 3
Chapter 3

Advertisement

Table of Contents
loading

Table of Contents