Model 3060E
Model 3060E
Model 3060E
Model 3060E
Model 3060E
HIGH CURRENT
FAILURE
(>30ppm)
CHECK FOR
AIR LEAKS
IN SAMPLE
LINE
PRESS START KEY
TO RESET THE ANALYZER
TO FACTORY
SETTINGS
FAILURE
CORRECTED
YES
PROCEED
5-12
CHECK SENSOR
1
MOUNTS
NO
CHECK BYPASS VENT
2
IS SET TO ~ 1 L/MIN
3
SET SPARGER TO
MEDIUM
O-RINGS
PROPERLY INSTALLED
MOUNTING
BOLTS
TIGHT
YES
PRESS START KEY
TO RESET THE ANALYZER
TO FACTORY
SETTINGS
YES
FAILURE
CORRECTED
NO
REPLACE
ELECTROLYTE
YES
FAILURE
CORRECTED
Table 5-8. High Current Failure
Teledyne Electronic Technologies
Analytical Instruments
PERFORM
NO
INDICATED
SERVICE
(INSTALLATION)
NO
CALL
SERVICE
Chapter 5
Chapter 5
Chapter 5
Chapter 5
Chapter 5