Chapter 5 Error Messages

Calibration Errors

Calibration Errors
The following errors indicate failures that may occur during a calibration.
Refer to the Service Guide for more information.
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside
the meter. When applicable, this error will occur at power-on to warn you
that the meter is unsecured.
702
Cal secured
The meter is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the meter. You must use the same security code to
unsecure the meter as was used to secure it, and vice versa. The security
code may contain up to 11 alphanumeric characters. The first character
must be a letter.
704
Secure code too long
A security code was received which contained more than 11 characters.
705
Cal aborted
A calibration in progress is aborted when you press any front-panel key,
send a device clear, or change the local/remote state of the meter.
Cal value out of range
706
The specified calibration value (CAL:VALue) is invalid for the present
function and range.
707
Cal signal measurement out of range
The specified calibration value (CAL:VALue) does not match the signal
applied to the meter.
709
No cal for this function or range
Attempt to calibrate gain on 1 mV range.
710
Full scale correction out of range
210