Chapter 3 Features and Functions
Triggering

Automatic Trigger Delays

If you do not specify a trigger delay, the meter selects an automatic delay
for you. The delay is determined by function, range, and integration time.
DC voltage :
Range
1 m V
10 m V to 120 V
Resistance (2-wire and 4-wire):
Range
1
10
100
1 k
10 k
100 k
1 M
Temperature:
RTD
THERmistor
ThermoCouple
<
NPLC
1
NPLC
Trigger Delay
Trigger Delay
15 ms
15 ms
1.0 ms
1.5 ms
NPLC < 1
NPLC
Trigger Delay
Trigger Delay
1.0 ms
1.5 ms
1.0 ms
1.5 ms
1.0 ms
1.5 ms
1.0 ms
1.5 ms
1.0 ms
1.5 ms
4 ms
6 ms
40 ms
60 ms
NPLC < 1
NPLC
Trigger Delay
Trigger Delay
1.0 ms
1.5 ms
1.0 ms
1.5 ms
1.0 ms
1.5 ms
³
1
³
1
³
1