Chapter 3 Features and Functions

Resistance Measurement Configuration

Resistance Measurement Configuration
The meter can make 2-wire or 4-wire ohms measurements. The meter
can compensate for voltages in the resistive circuit being measured using
offset compensation. You can also choose to either limit the power applied
or limit the open circuit voltage applied during a resistance
measurement.
The test current sourced by the meter depends upon the range and
conditions as shown below:
Range
2-wire or 4-wire
measurements
1
10
100
1 k
10 k
100 k *
1 M *
* Offset Compensation is not applied to measurements on these ranges
68
Test current
4-wire measurements
Low Power
10 mA
10 mA
10 mA
10 mA
10 mA
1 mA
1 mA
100 A
100 A
10 A
10 A
5 A
5 A
5 A
Voltage Limited
Not Available
1 mA
100 A
Not Available
Not Available
Not Available
Not Available