Agilent Technologies 34420A User Manual page 137

Nano volt/micro ohm meter
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
MEASure[:VOLTage][:DC]:DIFFerence?
[{< range > | AUTO | MIN | MAX | DEF}][,{< resolution > | MIN | MAX | DEF}]
This command presets and makes a difference measurement with the
specified range and resolution. The reading is sent to the output buffer.
For difference measurements, the specified range applies to the signal
connected to the Channel 1 terminals. Autoranging is selected for voltage
measurements on the Channel 2 terminals.
MEASure:FRESistance?
MEASure:RESistance?
[{< range > | AUTO | MIN | MAX | DEF}][,{< resolution > | MIN | MAX | DEF}]
This command presets and makes a 2-wire or 4-wire ohms measurement
with the specified range and resolution. The reading is sent to the output
buffer.
MEASure:TEMPerature?
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[,1,{< resolution > | MIN | MAX | DEF}]
{THER | DEF},{DEF}[,1,{< resolution > | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[,1,{< resolution > | MIN | MAX | DEF}]
This command presets and configures the meter for temperature
measurements using the specified probe type and resolution. The reading
is sent to the output buffer. DEF defaults to the last probe type set.
CONFigure[:VOLTage][:DC]
[{< range > | AUTO | MIN | MAX | DEF}][,{< resolution > | MIN | MAX | DEF}][,(@< channel >)]
This command presets and configures the meter for dc voltage
measurements with the specified range, resolution, and input channel.
This command does not initiate the measurement.
CONFigure[:VOLTage][:DC]:RATio
[{< range > | AUTO | MIN | MAX | DEF}][,{< resolution > | MIN | MAX | DEF}]
This command presets and configures the meter for dc:dc ratio
measurements with the specified range and resolution. This command
does not initiate the measurement. For ratio measurements, the specified
range applies to the signal connected to the Channel 1 terminals.
Autoranging is selected for voltage measurements on the Channel 2
terminals.