Calibration Count; Calibration Errors - Agilent Technologies U1241B User's And Service Manual

Handheld digital multimeters
Hide thumbs Also See for U1241B:
Table of Contents

Advertisement

5
Performance Tests and Calibration

Calibration Count

The meter provides the calibration count information for users to access
through front panel operation. Note that the meter was calibrated prior to
shipping out to users. Users are recommended to record the initial value
of the calibration count once receiving the meter.
The count value increases by one for each calibration point, from 0000 up
to the maximum of 19999. After the maximum count, the calibration count
will be reset to 0. The calibration count can be read from the front panel
after the meter has been unsecured, see the following steps:
1 In adjustment mode, press and hold
2 Take note of the calibration count to keep track of the number of
3 Press and hold

Calibration Errors

The following error codes indicate failures that may occur during a
calibration. The error code is displayed on secondary display.
Table 5-5 Calibration error codes
Code
56
view the calibration count viewing mode. The primary display indicates
the calibration count value while the secondary display indicates "Cnt".
calibration counts that has been performed.
count mode.
Descriptions
200
Calibration error: Calibration mode is secured
E02
Calibration error: Invalid secure code
E03
Calibration error: Invalid serial number code
E04
Calibration error: Calibration aborted
E05
Calibration error: Value out of range
E06
Calibration error: Signal measurement out of range
E07
Calibration error: Frequency out of range
E08
EEPROM write failure
for more than one second to exit the calibration
U1241B/U1242B User's and Service Guide
for more than one second to