Chapter 4 Remote Interface Reference
The SCPI Status Model
The Questionable Data Register
The questionable data register provides information about the quality of the
meter's measurement results. Overload conditions can be reported in the
questionable data summary bit through the enable register. You must write a
decimal value using the STATus:QUEStionable:ENABle command to set
the enable register mask.
Note: A reading overload condition is always reported in both the
standard event register (bit 3) and the questionable data event register
(bits 0, 1, or 9). However, no error message is recorded in the meter's error
queue.
Bit Definitions — Questionable Data Register
Bit
Voltage Overload
0
1
Not Used
2
Not Used
3
Not Used
Temperature Overload
4
5
Not Used
6
Not Used
7
Not Used
8
Not Used
Ohms Overload
9
10
Not Used
11
Not Used
12
Not Used
13
Not Used
14
Not Used
15
Not Used
Decimal
Definition
Value
1
Range overload on dc volts
2
Always set to 0.
4
Always set to 0.
8
Always set to 0.
16
Range overload on temperature
measurements.
32
Always set to 0.
64
Always set to 0.
128
Always set to 0.
256
Always set to 0.
512
Range overload on 2-wire or 4-wire
Ohms
1024
Always set to 0.
2048
Always set to 0.
4096
Always set to 0.
8192
Always set to 0.
16384
Always set to 0.
32768
Always set to 0.