Using The Range And Resolution Parameters - Agilent Technologies 34420A User Manual

Nano volt/micro ohm meter
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
Simplified Programming Sequence

Using the range and resolution Parameters

With the MEASure? and CONFigure commands, you can select the
measurement function, range, and resolution all in one command. Use
the range parameter to specify a fixed range larger than the expected
value of the input signal. You can set the range parameter to AUTO to
set autoranging.
Use the resolution parameter to specify the desired resolution for the
measurement. Setting the resolution sets the integration time of the
meter. The resolution parameter can be set to any number but the
resulting integration time (NPLC) is one of the following:
< resolution >
Parameter
< 0.0001 x Range
< 0.00001 x Range
< 0.000003 x Range
< 0.0000022 x Range
< 0.000001 x Range
< 0.0000008 x Range
< 0.0000003 x Range
< 0.00000022 x Range
Specify the resolution in the same units as the measurement function,
not in number of digits. For example, for volts measurements, specify the
resolution in volts. For resistance, specify the resolution in ohms.
You must specify a range to use the resolution parameter.
Integration Time
(NPLC)
0.02
.2
1
2
10
20
100
200
4
129