Statistics Operation Commands (Average) - Agilent Technologies 34420A User Manual

Nano volt/micro ohm meter
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
Math Operation Commands

Statistics operation commands (AVERage)

You must have set CALC:FUNC AVER and CALC ON to use these
commands.
CALCulate:AVERage:MINimum?
This command returns the minimum value found during a math average
operation. The meter clears the value when math is turned on, when
power has been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:MAXimum?
This command returns the maximum value found during a math average
operation. The meter clears the value when math is turned on, when
power has been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:AVERage?
This command returns the average of all readings taken since math
average was enabled. The meter clears the value when math is turned
on, when power has been off, or after a remote interface reset.
[volatile memory]
CALCulate:AVERage:COUNt?
This command returns the number of readings taken since math average
was enabled. The meter clears the value when math is turned on, when
power has been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:PTPeak?
This command returns the peak-to-peak value of all readings taken since
math was enabled. The meter clears the value when math is turned on,
when power has been off, or after a remote interface reset.
[volatile memory]
CALulate:AVERage:SDEViation
This command returns the standard deviation of all readings taken since
math was enabled. The meter clears the value when math is turned on,
when power has been off, or after a remote interface reset.
[volatile memory]