Chapter 3 Features and Functions

Defaults, Power-On and Reset States

Defaults, Power-On and Reset States
The meter stores settings in either volatile or non-volatile memory.
Settings stored in volatile memory are returned to default settings at
power-on or after a remote reset. Settings stored in non-volatile memory
are not changed by power-on or a remote reset (*RST), CONFigure, or
MEASure command.
Volatile memory settings and their defaults:
Setting
Function
Input Channel
Digital Filter
Analog Filter
Offset Compensation
N Samples
Display (On or Off)
Trigger
Math
Readings Hold
Readings Store
Trigger Delay
Integration Time (NPLC)
3
Null
3
Range
1
Remote defaults are used following reset (*RST), preset (SYSTem:PRESet), CONFigure,
or MEASure commands.
2
You should ensure the digital filter is turned off when using the remote interface. See page 57.
3
Integration Time, Null, and Range are independent in each function. CONFigure or MEASure
commands will only affect the Integration Time, Null, or Range for the function used.
112
Front Panel Default
DCV
Channel 1
On (50 rdgs)
Off
On
1
On
Autotrigger
Off
Off
Off
Auto
3
(10 plc)
Off
Autorange
1
Remote Default
DCV
Channel 1
2
Off
Off
Off
1
On
Autotrigger
Off
Off
Off
Auto
(10 plc)
Off
Autorange