Agilent Technologies 34420A User Manual page 149

Nano volt/micro ohm meter
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
Input Filter Commands
INPut:FILTer
:STATe {OFF | ON}
:STATe?
These commands enable or disable the filter state and query the filter
state.
INPut:FILTer
:TYPE {ANALog | DIGital | BOTH}
:TYPE?
These commands set the filter type and query the filter type.
BOTH enables the digital and analog filters.
INPut:FILTer
:DIGital:RESPonse {SLOW | MEDium | FAST}
:DIGital:RESPonse?
These commands set the averaging for the digital filter and query the
digital filter setting. SLOW sets a 100 reading average, MEDium sets a
50 reading average, and FAST sets a 10 reading average.
INPut:FILTer
:DIGital:PRECharge {OFF | ON}
:DIGital:PRECharge?
These commands enable or disable the digital filter precharge and query
the digital filter precharge setting. Disabling the digital filter precharge
prevents the filter from resetting based upon the measured value (see
page 57 in chapter 3).
4
147

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents