Math Operations
Statistics
After you enable statistics, the first reading that the meter takes is
stored as both the minimum and maximum value. The minimum is
replaced with any subsequent value that is less. The maximum is
replaced with any subsequent value that is greater.
All values are stored in volatile memory; the meter clears the values
when stats is turned on, when power has been off, or after a remote
interface reset or preset.
When enabled, the following statistics are available:
MIN
MAX
AVER
STD DEV
PEAK – PEAK peak to peak value of minimum and maximum
COUNT
Front Panel operation: Enable the statistics by pressing
Stats
the
key. Read the statistics gathered in the MATH menu
under the STATS command.
After reading the stored statistics, press
return to the measurement and continue taking statistics.
minimum reading
maximum reading
arithmetic average of all readings
standard deviation of all readings
the total number of readings
Shift
<
(Menu On/Off) to
75